There was no death or device malfunction associated with the defibrillation event.There is no information to suggest the patient sustained a serious injury.Device evaluation summary: monitor sn (b)(4) and belt sn (b)(4) were returned and evaluated at the distributor, in accordance with procedures recommended by zoll manufacturing corporation.The evaluation included review of downloaded software flag files on the day of the event and incoming functional testing.The review of the software flags consisted of an analysis of the downloaded data to identify any fault flags or unusual patterns of software flags.The software flag files did not suggest a device malfunction that would contribute to the inappropriate treatment event.During the incoming functional testing, a 1 hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5 hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5 hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no indication of a product malfunction contributing to the defibrillation event.The investigation into the event concludes that there was no device malfunction.A cause and effect analysis was conducted using all of the available information which includes the incident report, device evaluation, software flag files, and ecg strips (attached).The primary cause of the inappropriate shock event was lack of response button use prior to shock delivery.The ecg analysis, conducted by trained ecg technicians, identified the primary cause of the false detection as atrial fibrillation exceeding the programmed rate threshold.The rapid rate satisfied the rate detector of the detection algorithm.A secondary cause was motion artifact.The source of the motion artifact could not be positively identified through the cause and effect analysis.The following factors could not be ruled out as contributing causes of the motion artifact: body motion, poor ecg contact with skin.Inappropriate defibrillations are an anticipated risk associated with the use of the lifevest.Patients are instructed through alarms, voice messages, ifu, and training to press the response buttons to prevent an inappropriate defibrillation.(b)(4).A summary of the safety and effectiveness data (ssed), including the inappropriate defibrillation safety objective supporting fda's approval of the lifevest, can be found at http://www.Accessdata.Fda.Gov/cdrh_docs/pdf/p010030b.Pdf.The lifevest detection algorithm complies with iec (b)(4) performance requirements for sensitivity and specificity.
|