Model Number CD3371-40QC
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Patient Involvement (2645)
Event Date 06/23/2018
Event Type  malfunction  
Event Description
Upon interrogation, the device was observed to be in back-up vvi mode in the box.The issue was resolved by replacing the device.No patient was involved.
 
Manufacturer Narrative
The reported event of reset was confirmed.The device image was analyzed and the device memory was found to be corrupt.The device memory was tested in the laboratory and no anomaly was found.The memory corruption was not reproduced.As no data could be extracted from the device, the cause of the report event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7678913
MDR Text Key113704824
Report Number2938836-2018-08394
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model NumberCD3371-40QC
Device Lot NumberA000050602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-