• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER P.I. MUSCULAR VSD OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER P.I. MUSCULAR VSD OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-VSDMUSCPI-020
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative
An event of embolization was reported.The results of the investigation are inconclusive since the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
A 20 mm amplatzer p.I.Muscular vsd occluder was implanted on (b)(6) 2018 for treatment of a post-infarct vsd.After the procedure, anesthesia was reversed and the patient was noted to be disorientated and aggressively moving around.It was then noted that the occluder had embolized into the left pulmonary artery.On (b)(6) 2018, the occluder was surgically removed from the left pulmonary artery and patch closure of the vsd was performed.On a pre-surgery, ct, the patient was noted to have a pericardial hematoma, and the surgeon was able to suture repair the lpa where the occluder had been sitting.The device was discarded at the hospital and will not be returned for analysis.The patient is reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER P.I. MUSCULAR VSD OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7679455
MDR Text Key113742456
Report Number2135147-2018-00087
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model Number9-VSDMUSCPI-020
Device Catalogue Number9-VSDMUSCPI-020
Device Lot Number5722869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/15/2018
Initial Date FDA Received07/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-