Model Number M00564820
Device Problems Break (1069); Positioning Problem (3009); Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/20/2018
Event Type  Injury  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).The complainant indicated that the device has been contaminated and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2018 that an ultraflex tracheobronchial covered distal release stent was to be used to treat a malignant stricture in the airway during a stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was tight.According to the complainant, during the procedure, the stent was able to be deployed however, the proximal end of the stent was not fully expanded.The physician tried to pull the stent retention suture to reposition the stent, but the stent suture broke and the stent could not be adjusted.The stent was removed with forceps and another ultraflex tracheobronchial stent was placed to complete the procedure.There were no patient's complication reported as a result of this event.The patient's condition at the end of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX¿ TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7682985
MDR Text Key113815532
Report Number3005099803-2018-02299
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/10/2018
Device Model NumberM00564820
Device Catalogue Number6482
Device Lot Number0019948852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/20/2018
Initial Date FDA Received07/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-