The external visual inspection revealed the electrode was severed near the array and tool marks were observed on the top and bottom covers prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection confirmed severed electrode wires.This is believed to have occurred during revision surgery.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The scanning electron microscopy analysis revealed tool marks on the seam weld.The failure of this device is attributed to a short from the power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on the case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.In addition, this device had nitrogen that exceeded the limit.Based on an assessment of the residual gas analysis test data and feedthru dye penetrant inspection, it is determined that this device was non-hermetic, and that the root cause of the excessive nitrogen was a leak through the feedthru seals.Due to the presence of moisture getter, no signs of moisture were observed.This is the final report.
|