C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL 25BX; STATLOCK Back to Search Results
Catalog Number FOL0101
Device Problems Use of Device Problem (1670); Fitting Problem (2183)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the statlock device the patient attempted to use did to fit with the catheter from tray 900916a.The statlock was not apart of the tray.Per medical services, fol0101 is used for a 2 way foley catheter.The catheter in product 900916a is a temp sensing foley catheter that requires statlock item fol0105 which is designed for three way catheters.
 
Manufacturer Narrative
The reported event was confirmed use-related.No sample was returned for evaluation.Based on the initial investigation findings, current manufacturing controls are considered adequate as to detect and segregate any nonconforming unit.Event described could not be confirmed as a manufacturing related issue.The ifu currently addresses potential risks.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "statlock® foley stabilization device read carefully before use.Safety and efficacy considerations: single use.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterilized using ethylene oxide.Sterile unless package is opened or damaged, except for any individually packaged components within the pouch which are not labeled as sterile.These components are not terminally sterilized.Do not alter the statlock® device or components.Procedure must be performed by trained personnel with knowledge of anatomical landmarks, safe technique and potential complications.Contents: package includes the statlock® device stabilization system and skin preparation pad.Indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control rocedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.Technique prep 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel 7.Align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique disengage 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove.".
 
Event Description
It was reported that the statlock device the patient attempted to use did to fit with the catheter from tray 900916a.The statlock was not apart of the tray.Per medical services, fol0101 is used for a 2 way foley catheter.The catheter in product 900916a is a temp sensing foley catheter that requires statlock item fol0105 which is designed for three way catheters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL 25BX
Type of Device
STATLOCK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7689742
MDR Text Key114411649
Report Number1018233-2018-02786
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076091
UDI-Public(01)00801741076091
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFOL0101
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/25/2018
Initial Date FDA Received07/16/2018
Supplement Dates Manufacturer Received07/31/2018
Supplement Dates FDA Received08/23/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-