Model Number RSINT35015X |
Device Problems
Material Deformation (2976); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
During the procedure, an attempt was made to use one resolute integrity drug eluting stent to treat a severely tortuous and moderately calcified lesion in the lad.It was reported that the was damaged during use.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues identified.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.The lesion was too calcified, and the device would not pass through the lesion.Patient status post-procedure is alive with no injury.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|