MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160177
Device Problems Crack (1135); Material Deformation (2976); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/20/2018
Event Type  malfunction  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
As reported: "the plastic blue piece that goes on the tibial c impactor is cracked and falls off during the case".
 
Manufacturer Narrative
An event regarding crack/fracture involving an mako impactor was reported.The event was confirmed through product return.The blue foot attachment for the tibial impactor was returned for evaluation.The device was returned in a used condition.Scratches and dent marks are observed on the surface of the device.Examination of the returned device with engineer indicated that the device is worn due to in service use consistent with contact against hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.No medical records were received for review with a clinical consultant all devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the reported lot.The blue foot attachment for the tibial impactor was returned for evaluation.The device was returned in a used condition.Scratches and dent marks are observed on the surface of the device.Examination of the returned device with engineer indicated that the device is worn due to in service use consistent with contact against hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
As reported: "the plastic blue piece that goes on the tibial c impactor is cracked and falls off during the case".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INLAY IMPACTOR HEAD
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key7699906
MDR Text Key114429115
Report Number3005985723-2018-00427
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160177
Device Lot Number12150617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Initial Date Manufacturer Received 06/20/2018
Initial Date FDA Received07/18/2018
Supplement Dates Manufacturer Received07/30/2018
Supplement Dates FDA Received08/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-