Catalog Number 2400DX
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 04/16/2018
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation and the evaluation is complete.A visual inspection was performed on the returned device and noted a punctured display cable with exposed wire.A service history review and device history review were performed and found no indications of a nonconformance during the manufacture of this product.The reported condition was verified.The cause of the condition could not be determined.The device was removed from service.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the display cord of a em2400 new display was found to be damaged and frayed with exposed wires.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM2400 NEW DISPLAY
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
ni
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION
ni
ni
ni
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7704813
MDR Text Key114554253
Report Number1416980-2018-04456
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400DX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2018
Initial Date FDA Received07/19/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-