The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome membrane leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot g317 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot g317 for the reported issue shows no trends.Trends were reviewed for complaint category, pressure dome membrane leak.No trends were detected for this complaint category.The kit and smartcard were returned for investigation.A review of the data recorded on the smartcard revealed a system pressure alarm occurred during the procedure.The received kit was examined and found the system pressure dome diaphragm partially unseated from the pressure dome housing.The unseated system pressure dome diaphragm was cleaned and reinstalled into the housing.The system pressure dome was installed onto a test pressure sensor and fit without any issues.A pressure test was performed while the pressure dome was installed on the sensor and no leaks were identified.The pressure dome membrane leak is confirmed based on the received kit and finding of the diaphragm partially unseated from the housing.The root cause of the leak is most likely improper installation of the pressure dome by the user.If the pressure dome is not properly seated and latched around the pressure sensor, the system pressure may force the diaphragm to detach from the housing resulting in a leak.No further action is required at this time.This investigation is now complete.(b)(4).(b)(6) 2018.
|