Model Number 9-PFO-025
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/22/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 25mm amplatzer pfo occluder was implanted.After releasing the device from the delivery cable, the device was observed on tee to have moved and was standing steep on the septum.The physician decided to leave the device implanted and re-evaluate later the same day.The follow-up tee showed the device had embolized to the right ventricle.The patient was transferred to cardiac surgery and the device was explanted and the defect closed.The patient is reported to be stable.
 
Manufacturer Narrative
An event of device embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7708806
MDR Text Key114670802
Report Number2135147-2018-00092
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806011288
UDI-Public00811806011288
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Patient Monitoring
Type of Report Initial,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date02/28/2023
Device Model Number9-PFO-025
Device Catalogue Number9-PFO-025
Device Lot Number6384772
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/22/2018
Initial Date FDA Received07/21/2018
Supplement Dates Manufacturer Received06/22/2018
Supplement Dates FDA Received08/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-