MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA1D4
Device Problems Nonstandard Device (1420); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Low Blood Pressure/ Hypotension (1914); Low Oxygen Saturation (2477); Diminished Pulse Pressure (2606)
Event Date 06/21/2018
Event Type  Death  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, the patient experienced worsening hypotension and oxygen saturation.The patient was then intubated and converted to general anesthesia.Ultimately the cardiac resynchronization therapy defibrillator (crt-d) system was implanted and the patient was transferred to the critical care unit.Shortly after transfer the patient became pulseless and was resuscitated with cardiopulmonary resuscitation, epinephrine, bicarbonate, calcium, norepinephrine, and cardioversion from ventricular fibrillation (vf) to atrial fibrillation (af).Ultimately the patient developed a sinus rhythm but they remained pulseless.An echocardiogram was performed and it was negative for a pericardial effusion.The decision was made to withdraw support and the patient died.A cause of death of acidemia with hypotension and electrolyte imbalance underlying causes of dilated cardiomyopathy with an ejection fraction of ten percent and chronic kidney disease with acute injury was provided.The patient's physician also noted that they felt the patient was so compromised that they were unable to tolerate the anesthesia required for any procedure.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7710293
MDR Text Key114705626
Report Number3004209178-2018-16370
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169720657
UDI-Public00643169720657
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/14/2019
Device Model NumberDTMA1D4
Device Catalogue NumberDTMA1D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2018
Initial Date FDA Received07/23/2018
Supplement Dates Manufacturer Received07/31/2018
08/08/2018
08/08/2018
Supplement Dates FDA Received08/01/2018
10/04/2018
10/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0962-2017
Patient Sequence Number1
Treatment
6935M62 LEAD, 5076-52 LEAD
Patient Outcome(s) Death;
Patient Age62 YR
-
-