• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. PALINDROME; CATHETER, HEMODIALYSIS, IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. PALINDROME; CATHETER, HEMODIALYSIS, IMPLANTED Back to Search Results
Model Number 8888145039P
Device Problems Break (1069); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/06/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device was replaced due to a blood leak.It was stated that the device was placed.There was no patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: that the distal end of the catheter was cut.The plug caps and dual extension appeared intact.A functional evaluation found that submerging the catheter into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Both extensions were tested with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Type of Device
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer Contact
lisa hernandez
15 hampshire street
mansfield, MA 02048
2034925563
MDR Report Key7713689
MDR Text Key115063789
Report Number3009211636-2018-00246
Device Sequence Number1
Product Code MSD
UDI-Device Identifier10884521158023
UDI-Public10884521158023
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123196
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/22/2020
Device Model Number8888145039P
Device Catalogue Number8888145039P
Device Lot Number1602100116
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/09/2018
Initial Date FDA Received07/24/2018
Supplement Dates Manufacturer Received08/08/2018
09/18/2018
Supplement Dates FDA Received10/03/2018
10/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-