• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-018
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2018
Event Type  Injury  
Manufacturer Narrative
An event of device embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2018, a pfo was measured to be 14mm on echocardiogram and a 25mm amplatzer cribriform (lot number: 6469071) was selected for implant.During a tug test, the device was deemed too small and the device was re-sheathed and removed from the patient.A 24mm sizing balloon was used to re-size the defect, measuring 18mm, and an 18mm amplatzer septal occluder (aso, lot number: 6165706) was successfully implanted after checking stability.No leaks were observed and the device was released.Later that day, the 18mm aso embolized to the aorta.The device was percutaneously snared and retrieved from the patient.There were no patient consequences, and pfo closure will be re-evaluated by the physician on a later date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7716014
MDR Text Key114934135
Report Number2135147-2018-00095
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010151
UDI-Public00811806010151
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Repair
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number9-ASD-018
Device Catalogue Number9-ASD-018
Device Lot Number6165706
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/25/2018
Initial Date FDA Received07/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-