(b)(4).A field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse confirmed the reported issue via the error log.The fse replaced the sv3 then verified the sample results, precision, and quality control (qc).All results were within acceptable range.All errors cleared and the g8 instrument was operational.No further actions were required by field service.The 2-way solenoid valve was returned for investigation to the instrument service center (isc).Functional testing of the returned part confirmed the reported error.The valve seat was not working.A 13-month complaint / service history review for similar complaints was performed for the serial number (b)(4) from aware date of (b)(6) 2018, which included data from 27 may 2017 through 27 june 2018.There was one similar complaint identified during the searched period, which includes this event.The g8 operator's manual under chapter 6: troubleshooting is as follows: 101 pressure low the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause for the reported event was the faulty sv3 2-way solenoid valve.
|
A customer reported error 101 pressure low on the g8 instrument.The customer changed the filter, column, and buffers.The customer ran multiple specimens without error.While the g8 was in the wash cycle after the run was completed, the low pressure error appeared again.The customer informed the technical support specialist (tss) that there was no evidence of a fluid leak.The customer is unable to run patient samples on hba1c.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|