(b)(4).A field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse found the buffer valve # 3 was faulty and buffer valve # 1 was leaking.The fse replaced both valves.All errors cleared and the g8 instrument was operational.No further actions were required by field service.A 13-month complaint/service history review for similar complaints was performed for the serial number (b)(4) from aware date of 29 june 2018, which included data from 29 may 2017 through 29 june 2018.There was one similar complaint identified during the searched period, which includes this event.The g8 operator's manual under chapter 6 troubleshooting is as follows: the 101 pressure low: the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause for the reported event was the defective buffer valve.
|