The device was received for evaluation.Visual inspection revealed that the power/communication cable insulation was punctured, exposing the ground shielding.The internal wires were not exposed.The reported condition of a frayed cable was verified.The cause of the condition was determined to be that the display cable was not properly routed through the cut-out slot of the main module cover, which led to the tubing being punctured from the weight of the main module sitting on the cable.The display was removed from service.Should additional relevant information become available, a supplemental report will be submitted.
|