SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TAH-T; BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Catalog Number 500101
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Neurological Deficit/Dysfunction (1982); Renal Failure (2041); Urinary Tract Infection (2120); Fungal Infection (2419); Respiratory Failure (2484)
Event Type  Injury  
Manufacturer Narrative
The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.Investigation summary: intermacs patient registry data collected between april 1, 2014 and september 30, 2017 regarding adverse events was retrospectively reviewed.Patients were de-identified therefore a one-to-one correlation could not be made between the patient and the lot number of the implanted tah-t.These events were assessed and are being reported in response to an update to the mdr complaint sources.With a review of the available information, there is no evidence of a device malfunction or performance issues that would impact the reported events.Possible clinical factors that may have contributed to these events include the patients pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulants, antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to these events.This device is used for treatment, not diagnosis.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4).(b)(6).
 
Event Description
While supported by the tah-t, the patient experienced the following adverse events as defined by intermacs: 8 days post implant - neurological dysfunction - cva/location: left hemisphere: occipital, left hemisphere: parietal/severity: other; 10 days post implant - renal dysfunction; 10 days post implant - respiratory failure; 1 day post implant - infection/location: mediastinum/type: bacterial; 16 days post implant - infection/location: pulmonary/type: fungal; 34 days post implant - infection/location: urinary tract/type: bacterial; 62 days post implant - neurological dysfunction - cva/location: right hemisphere: unspecified/severity: left sided weakness.The patient subsequently received a heart transplant after 76 days of tah-t support.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TAH-T
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7726547
MDR Text Key115236307
Report Number3003761017-2018-00362
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Remedial Action Inspection
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number500101
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/28/2017
Initial Date FDA Received07/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight73
-
-