BECKMAN COULTER COULTER AC·T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number AC*T DIFF 2
Device Problem Output Problem (3005)
Patient Problem Test Result (2695)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
On 10 jan 2017, the field service engineer (fse) was at the customer site and confirmed that plts were running high and mcvs were running low on the coulter ac·t diff 2 analyzer.The fse replaced the sweep flow check valve to resolve the reported event.The repairs were verified by the fse.The failure mode is attributed to the sweep flow check valve.Beckman coulter is conducting a retrospective review of its global service records, to ensure that the records had been adequately evaluated for potential complaint information and any warranted follow-up action, including mdr reporting.  this mdr report is a result of that retrospective review.  (b)(4).
 
Event Description
The customer reported that their coulter ac·t diff 2 analyzer generated erroneous high platelet (plt) results on patient samples.There was no report of erroneous results reported out of the lab from the provided information.There was no report of change or effect to patient treatment in connection with the reported event.Instrument printouts of patient results or controls were not provided for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC·T DIFF 2 ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
laurie o'riordan
11800 sw 147th avenue
miami, FL 33196-2031
3053802874
MDR Report Key7726907
MDR Text Key115361100
Report Number1061932-2018-01020
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier15099590264796
UDI-Public(01)15099590264796(11)NO-DATA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAC*T DIFF 2
Device Catalogue Number6605500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2009
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-