• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G148
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2018
Event Type  malfunction  
Manufacturer Narrative
At this time, the product remains in-service.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this patient with cardiac resynchronization therapy defibrillator (crt-d) and right atrial (ra) lead had high out of range pacing impedance greater that 2000 ohms, where the minute ventilation (mv) signal was being oversensed.The mv feature was turned off in response, and the plan was to continue to monitor.The device remains in-service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7727504
MDR Text Key115361036
Report Number2124215-2018-13694
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/07/2017
Device Model NumberG148
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/26/2018
Initial Date FDA Received07/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-