• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3228
Device Problem Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 03/05/2017
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient was not able to turn on her scs system's stimulation.The patient stated when she attempted to turn on her stimulation the system would go down to zero.Diagnostics revealed multiple lead contacts with invalid impedance.Reprogramming was performed but the issue persisted.Subsequently, surgical intervention was taken and the patient's scs lead was replaced.Effective stimulation therapy was reportedly restored postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
arnulfo ochoa
6901 preston road
plano, TX 75024
9723098090
MDR Report Key7733478
MDR Text Key115521469
Report Number1627487-2018-07077
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017253
UDI-Public05415067017253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Model Number3228
Device Lot Number5441304
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-