On (b)(6) 2018, the patient had a procedure with a residual pericardial effusion that the user attributes to that procedure.On (b)(6) 2018, a 24 mm amplatzer p.I.Muscular vsd occluder was selected for closure of a post-infarct vsd.A 1 mm lv pericardial effusion was noted initially on fluoroscopy at the start of the procedure.During the procedure, the patient became hemodynamically unstable and experienced hypotension at which time a large effusion was detected.The physician deployed the device in the 1 mm defect and did no further intervention for the remaining defects due to the patient condition.Despite efforts made to stabilize the patient; it is reported the patient died in the icu the following day.
|