Model Number 397002-001 |
Device Problems
Display or Visual Feedback Problem (1184); Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/18/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the issue was observed while the driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited diastolic waveform delay during a training session.
|
|
Manufacturer Narrative
|
The customer-reported waveform timing offset was reproduced during investigation testing in that a timing delay manifested in the right flow waveform.The root cause of the waveform timing offset was determined to be a malfunction of the right pilot valve, which caused a right flow waveform delay/offset.Syncardia has a corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation and is closing this file.(b)(4).
|
|