• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problem Computer Software Problem (1112)
Patient Problem No Patient Involvement (2645)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts have been returned to the manufacturer for analysis.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the mobile viewing station (mvs) boots well, but the image acquisition system (ias) was unresponsive and showing "initializing system." the fan never finished the initialization, so the site attempted to unplug it and it would still not go past the initializing screen.Site added that after unplugging and reconnecting the cable, the mvs was beeping loudly and did not go past the start windows screen.Site then restarted the system and ias was not going past the initializing screen and the mvs was beeping.Site was unable to "open image" and the fan loading on the scanner was not finished so the site couldn't do any type of scan.There was no patient present during this event.
 
Manufacturer Narrative
A software investigation analysis was initiated to determine the probable cause of the issue through review of the reported issue.Analysis found that the reported issue was unrelated to the software of the imaging system.Analysis found that the software functioned as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7737241
MDR Text Key115647873
Report Number1723170-2018-03775
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-