Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (constant gong alarms / does not latch) was confirmed.As received, the belt failed the therapy electrode recognition test and would not securely latch to a test monitor.Upon evaluation, there was an open pulse wire inside the cable connecting the distribution node (dn) and front therapy electrode (te), between the te and ecg electrode a.The cause of the constant gong alarms is the open wire.The root cause of the open wire is excessive force placed on the cable.Upon further evaluation, the tabs on the trunk cable connector were broken.The cause of the inability to securely latch to a monitor is the broken tabs.The root cause of the broken tabs is excessive force placed at the connector.No adverse event resulted from the defective electrode belt.
|