Model Number 01C-42640-06 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned to the manufacturer for investigation.It has not been received.Investigation is not complete.
|
|
Event Description
|
The event involved a transpac iv monitoring kit with safeset reservoir and blood sampling port that disconnected between the pressure line and the blood withdrawal port.The device was in use for approximately 1 hour with heparinized saline for blood pressure monitoring.The device was replaced with no further problems encountered.There was no adverse event or delay in critical therapy.
|
|
Manufacturer Narrative
|
Received one (1) used partial set 01c-42640-06 transpac iv monitoring kit with safeset, lot# 3514084 for testing.Testing found a tubing separation in the 27" arterial pressure tubing and the safeset port.There was sufficient solvent applied to the bond.A device history review (dhr) was completed and no non-conformances were noted.The reported complaint of tubing separation was confirmed.The cause of the separation appears to be due to excessive pulling force near the safeset port.
|
|
Search Alerts/Recalls
|