This report has been identified as b.Braun medical inc.Internal report number (b)(4).Apex technical support was able to confirm the issue of occlusions.The log review did not confirm any events for pump calibration failure.One (1) used set, without packaging, was returned for evaluation.Visual examination of the set noted no visual defects.The set was occlusion tested on line 1 with passing results.Visual examination of a retained transfer set noted no visual defects.The retained unit was occlusion tested with passing results.A review of our discrepancy management system database found no related or similar discrepancies during the production of the batch.Based on the results of the investigation this complaint is not confirmed for valve misalignment.B.Braun will continue to investigate, trend, and monitor all complaints of this nature.If additional pertinent information becomes available a follow-up report will be filed.
|