TELEFLEX MEDICAL HUDSON VOLDYNE2500 VOLUMETRIC EXERCISER; SPIROMETER, THERAPEUTIC (INCEN Back to Search Results
Catalog Number 8884719025
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device involved in this complaint has not been received by the manufacturer for evaluation at the time of this report.Verification of failure mode reported in the current manufacturing process was conducted.Thirteen device samples were taken from the current production.The samples were functionally inspected, and during testing, the issue reported was not observed in the current manufacturing process.A device history record investigation did not show issues related to this complaint.Customer complaint cannot be confirmed due the product sample is not available to perform a proper investigation and determine the root cause.Corrective actions cannot be established at this time.If the device sample becomes available at a later date, this report will be updated accordingly.
 
Event Description
Customer complaint alleges the yellow indicator peg sticks during use.Reported by a home care user.Customer states the device must be shaken to reset the peg.No harm was reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON VOLDYNE2500 VOLUMETRIC EXERCISER
Type of Device
SPIROMETER, THERAPEUTIC (INCEN
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
linda woodall
3015 carrington mill blvd
morrisville 27560
9196942566
MDR Report Key7753472
MDR Text Key116214381
Report Number3003898360-2018-00622
Device Sequence Number1
Product Code BWF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K801199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number8884719025
Device Lot Number01J1300525
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/25/2018
Initial Date FDA Received08/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-