Model Number 3228
Device Problem Impedance Problem (2950)
Patient Problems Device Overstimulation of Tissue (1991); Inadequate Pain Relief (2388)
Event Date 06/10/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient had been unable to receive adequate stimulation.Impedance issues were found on a majority of the patient's lead contacts.Reprogramming was ineffective long-term.Unspecified imaging was performed but the results remain unknown.It was also reported the patient had experienced overstimulation.As a result, the patient's lead was explanted and replaced.Surgical intervention addressed the patient's issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
drew johnson
9725264667
MDR Report Key7754784
MDR Text Key116199760
Report Number1627487-2018-07307
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017253
UDI-Public05415067017253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/13/2020
Device Model Number3228
Device Lot Number6403181
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/12/2018
Initial Date FDA Received08/06/2018
Supplement Dates Manufacturer Received08/17/2018
Supplement Dates FDA Received09/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-