Model Number G158
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2018
Event Type  Injury  
Manufacturer Narrative
The device remains in service at this time.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) exhibited right ventricular (rv) noise, oversensing, and pacing inhibition with a pause in pacing greater than three seconds.There was also noise on the shock channel.It was discussed that the noise appeared to be due to electromagnetic interference (emi).All lead measurements were stable.The sensitivity was reprogrammed and the patient would follow up with their physician.No additional adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7755710
MDR Text Key116204659
Report Number2124215-2018-14254
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/06/2018
Device Model NumberG158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 05/25/2018
Initial Date FDA Received08/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0292; 4675; 7741; G158
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-