MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G156
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2018
Event Type  malfunction  
Manufacturer Narrative
The local area field representative was contacted for additional information.It was reported the patient was brought into clinic for evaluation.Noise was unable to be recreated and impedance measurements remain stable.The respiratory sensor was programmed off to mitigate further oversensing on the atrial channel.The system will continue to be monitored.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d), implanted with another manufacturer's leads, recorded inappropriate atrial tachy response episodes due to oversensing of the respiratory sensor.Baseline noise was also noted on the right ventricular channel which appeared to be myopotential.This noise was not being sensed by the device.A spike in the right ventricular pacing impedances had been observed.Boston scientific technical services (ts) discussed possible causes and provided troubleshooting options.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7755802
MDR Text Key116221041
Report Number2124215-2018-14742
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/28/2018
Device Model NumberG156
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/07/2018
Initial Date FDA Received08/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4674; 5076; 6947; D154AWG; G156
Patient Age69 YR
-
-