U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G158
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited noise and oversensing due to electromagnetic interference (emi) during an unrelated ablation procedure.The noise and oversensing resulted in pacing inhibition and inappropriate anti-tachycardia pacing (atp) therapy being delivered.The crt-d remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7756131
MDR Text Key116219107
Report Number2124215-2018-15054
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public(01)00802526534904(17)20160514
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/14/2016
Device Model NumberG158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/28/2018
Initial Date FDA Received08/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076; G158; MISMATCH
Patient Age90 YR
-
-