• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 280025MM
Device Problems Device Dislodged or Dislocated (2923); Insufficient Information (3190)
Patient Problems Aortic Valve Stenosis (1717); Stenosis (2263)
Event Date 07/17/2018
Event Type  Injury  
Manufacturer Narrative
Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Svd, a common reason for bioprosthesis explant or reoperation, encompasses multiple failure modes, including calcific and non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.The device was not returned for evaluation, as it remained implanted.Based on the available information, the root cause of the event was likely due to patient related factors and the progression of the patient's underlying valvular disease pathology.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a model 25mm pericardial aortic valve was disabled via a valve-in-valve procedure after an implant duration of 11 years, 5 months due to stenosis.A 26mm transcatheter valve was implanted successfully.The patient tolerated the procedure well and was transferred to the icu in stable condition.The patient was discharged home in stable condition on post-operative day two.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Type of Device
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key7768799
MDR Text Key116626005
Report Number2015691-2018-03287
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
PMA/PMN Number
P860057/S001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2010
Device Model Number280025MM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/17/2018
Initial Date FDA Received08/09/2018
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received11/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age78 YR
-
-