Model Number 2000E |
Device Problem
Restricted Flow rate (1248)
|
Patient Problem
Death (1802)
|
Event Date 07/16/2018 |
Event Type
Death
|
Manufacturer Narrative
|
Concomitant medical products: d-60hl rapid infuser tubing, continu-flow solution set, 10 drops/ml product number: 2c8541, therapy date: (b)(6) 2018.No product will be returned per customer.The customer complaint could not be confirmed because the product was discarded and not returned for failure investigation.The root cause of this failure was not identified.
|
|
Event Description
|
The customer reported that the patient needed the use of pressure bags and a level 1 rapid infuser for multiple infusions and transfusions through multiple central lines.The users noticed impeded flow during the infusions.The rapid infuser and the tubing that appeared to be malfunctioning were replaced and the customer ensured that each iv port was flowing adequately, however the decreased flow rate continued.After placement of another large bore central line, the flow issue persisted.The physician felt that the needle free valve did not tolerate the high pressure required and may not have been compatible with the infuser tubing.A non-bd needlefree valve was used instead with no issues.The patient was further resuscitated, was later placed on comfort care, and passed away.An autopsy was declined by the patient's family and attending physicians.The customer does not believe the product issue caused the patient's death.
|
|