MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL INFUSION PUMP Back to Search Results
Model Number INFKIT2
Device Problem Failure to Auto Stop (2938)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
This device was returned to mmdg for evaluation.The pump operated within product specifications.The pump history was also reviewed where the pump shows that it had alarmed no flow out multiple times before it was turned off.Mmdg could not confirm this complaint.There is no indication that the complaint occurred as reported.This report is being filed because the event occurred while the device was in use by a pediatric patient.Per our procedures all reports of this type of event that involve a pediatric patient are considered reportable.
 
Event Description
The initial reporter stated that the pump was not alarming when occluded or when there was air in the line.Mmdg followed up with the initial reporter, who stated that the patient was stable, but they did not have any additional information regarding the complaint.[(b)(4)].
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL INFUSION PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin hardesty
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key7772989
MDR Text Key116875240
Report Number1722139-2018-00205
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Date Manufacturer Received07/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age3 YR
Patient Weight17
-
-