BOSTON SCIENTIFIC - MINN FG GATEWAY MR JP 3.50MM X 12MM; CATHETER, BALLOON TYPE Back to Search Results
Catalog Number M0032072112350
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 06/20/2018
Event Type  Injury  
Manufacturer Narrative
This is the fourth of 4 reports.Subject device is not available.
 
Event Description
Patient successfully underwent intracranial stenting with stent microcatheter and balloon catheters (subject device) on the internal carotid artery (ic) with stenosis rate of 80% and lesion length of 10 mm.Next day magnetic resonance imaging (mri t2) weighted image showed unknown black void on right brain.However, the same level of ct image showed no sign.According to the physician, the black void indicates metal artifact of some metal which was used in the procedure, possibly fragments of coating or a part of product components.The patient was asymptomatic.
 
Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, physical as well as a functional testing could not be performed.While there are a number of potential causes for the reported issue, because the review of all available information failed to identify a definitive cause and because the product was not returned for analysis, the exact cause for the reported un-retrieved device fragments cannot be determined.Expiration date: added.Manufacturing date: added.
 
Event Description
Patient successfully underwent intracranial stenting with stent microcatheter and balloon catheters (subject device) on the internal carotid artery (ic) with stenosis rate of 80% and lesion length of 10 mm.Next day magnetic resonance imaging (mri t2) weighted image showed unknown black void on right brain.However, the same level of ct image showed no sign.According to the physician, the black void indicates metal artifact of some metal which was used in the procedure, possibly fragments of coating or a part of product components.The patient was asymptomatic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FG GATEWAY MR JP 3.50MM X 12MM
Type of Device
CATHETER, BALLOON TYPE
Manufacturer (Section D)
BOSTON SCIENTIFIC - MINN
one scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC - MINN
one scimed place
maple grove MN 55311
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key7775639
MDR Text Key116856796
Report Number3008853977-2018-00040
Device Sequence Number1
Product Code GBA
Combination Product (y/n)N
PMA/PMN Number
H050001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/22/2019
Device Catalogue NumberM0032072112350
Device Lot Number20101486
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2018
Initial Date FDA Received08/12/2018
Supplement Dates Manufacturer Received08/14/2018
Supplement Dates FDA Received08/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age51 YR
-
-