Catalog Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one approved temporary deviation notice (dn) reported on the lot which was unrelated to the complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
A user facility nurse reported a blood leak that occurred 30 minutes into the patient¿s hemodialysis (hd) treatment.The blood leak was visually observed at the connection between the bloodlines and the blood pump (venous side).The patient¿s estimated blood loss (ebl) was approximately 100 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on the same machine, and completed treatment with new supplies with no harm, adverse event, or medical intervention required.The complaint device was discarded and is not available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7777360
MDR Text Key117020332
Report Number8030665-2018-01221
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03-2722-9
Device Lot Number18BR01122
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age MO
Initial Date Manufacturer Received 07/31/2018
Initial Date FDA Received08/13/2018
Supplement Dates Manufacturer Received08/13/2018
Supplement Dates FDA Received08/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
Patient Weight50
-
-