Event summary: visual inspection of catheter 2af283/ 06612 showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for eleven injections.Insertion and retraction tests were performed with difficulty.Inflation showed the kink on guide wire lumen under the balloon.Dissection showed a guide wire lumen kink at 1.41 inches from the tip inside the balloons.The catheter passed the performance test and electrical integrity as per specification, impedance was also within specification.In conclusion, the balloon catheter failed the inspection due to a guide wire lumen kink.No system notice triggered through testing for returned product.If information is provided in the future, a supplemental report will be issued.
|