(b)(4).Device evaluated by mfr: the device was returned for analysis.Returned product consisted of a direxion micro-catheter with the ¿y¿ adapter.The shaft was microscopically examined.The device showed 2 fractures on the nickel shaft.The 1st fracture was located 2mm from the hub.The 2nd fracture was located 37.5cm from the hub.No other damage was noticed on the shaft.The device was not completely separated the inner liner was intact.The ¿y¿ connector was returned with the device and applied to the manifold.A syringe with fluid was attached to the ¿y¿ connector and pressure was applied to the syringe.The fluid exited the tip of the device and no fluid was leaking from the hub or the ¿y¿ connector.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
Reportable based on device analysis completed on 24-jul-2018.It was reported that leak in the hub occurred.Vascular access was obtained via the right femoral artery.A direxion transend-14 system was selected for use.During procedure, a leakage from the y connector was noted.The procedure was completed with this device.No patient complications were reported.However, device analysis revealed a fracture on the nickel shaft 37.5cm from the hub.
|