Event summary: the patient data files showed at least four applications were performed with catheter 2af284 / 25282 on the date of the event with no issue.Additionally, the patient data files showed at least twenty applications were performed with a second catheter 2af284 / 25282 on the date of the event with no issue.Visual inspection of the balloon catheter showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for four injections.Dissection showed a guide wire lumen kink at 1.40 inches from the tip inside the balloons.Pressure test did not show leaks.In conclusion, the reported kink issue has been confirmed through testing.The balloon catheter failed the inspection due to guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the balloon catheter was kinked.The balloon catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.On 2018-07-24: the catheter was returned to the manufacturer, analyzed, and tested out of specification.
|