(b)(4).The sample was returned for evaluation.It was observed that the sample was received without the puncture pin protector.Functional testing was also performed and the sample passed the tests.During the setup of the oxygen entrainment testing, however, it was observed that the assembly of the nut adaptor and the upper body component was unstable.Even with that condition, the sample was able to be tested with no functional issues.After the testing was finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection it was found there was wear on the internal tabs.Attempts to duplicate the failure mode were performed and there are two ways to duplicate them: the first one is by overtightening the nut adaptor into the flow meter.The second one is by manipulating the assembly connection.Based on the investigation performed, the reported complaint was confirmed.Although the condition observed on the sample provided, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor most likely was caused by the end user during the connection of the adaptor into the flowmeter.However, the personnel of the assembly line were notified on (b)(6) 2018 for awareness.
|