MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE COVERED STENT Back to Search Results
Model Number LSMU1350837
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Death (1802)
Event Type  Death  
Manufacturer Narrative
The production history for the device was reviewed and no anomalies were identified.The lot met all test release criteria.Nothing was found to indicate a manufacturing related cause for this event.This is the first reported complaint for this lot number and issue to date.The result of the investigation is inconclusive.No sample was provided for evaluation.It is unknown if patient factors (the vessel been treated was highly calcified), procedural or handling techniques may have contributed to the reported event.The event description also states that the patient expired 3-4 hours post procedure.The cause of the patient¿s death was not provided and it is unknown if it was related to the device.Based upon the available information a definitive root cause has not been determined.The ifu for the lifestream product was reviewed and contains the following information relevant to the reported event: a.Device description: 1.Implant the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from (b)(4) stainless steel, encapsulated between two layers of eptfe.B.Indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation: 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection: 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation: 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.6.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation: 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.Expiry date: nov / 2020.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that during inflation of the pta balloon catheter to treat a lesion in the left iliac artery via a right common femoral approach, the balloon expandable covered stent allegedly dislocated from the balloon catheter.It was further reported that the health care provider exchanged the balloon catheter over the guide wire for another and was able to position and expand the covered stent in situ; however, the stent was not positioned at the lesion site.No further treatment was provided and the procedure was concluded.Reportedly, the patient expired approximately 3-4 hours post procedure.The cause of the patient¿s death was not provided.No other pertinent patient, device or medical information was provided leading up to or surrounding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE COVERED STENT
Type of Device
BALLOON EXPANDABLE COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI  
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key7793970
MDR Text Key117489104
Report Number9616666-2018-00112
Device Sequence Number1
Product Code PRL
UDI-Device Identifier05391522081222
UDI-Public(01)05391522081222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLSMU1350837
Device Catalogue NumberLSMU1350837
Device Lot NumberCMBZ0234
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
-
-