Model Number 25-092-12-91
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/26/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the screw broke during implantation.The screw was removed.
 
Manufacturer Narrative
An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number was not identified; therefore, the device history records were not reviewed.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Type of Device
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM   78570
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key7793987
MDR Text Key117503949
Report Number9610905-2018-00184
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00888118037220
UDI-Public(01)00888118037220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25-092-12-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/04/2018
Event Location Hospital
Date Report to Manufacturer07/26/2018
Initial Date Manufacturer Received 07/26/2018
Initial Date FDA Received08/17/2018
Supplement Dates Manufacturer Received10/04/2018
Supplement Dates FDA Received10/29/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
Patient Weight122
-
-