TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801040 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the user facility reported that they tried unplugging and plugging the cable back in several times, however the problem persisted.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there was poor communication between the roller pump and the central control monitor (ccm).As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the pump cable is broken but will not be returned for investigation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|