Catalog Number 05.001.201
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the mechanics and gear-wheel of the battery handpiece/modular device were damaged.It was further determined that the device ran hot.It was further determined that the device failed pretest for check function of all modes and check of free moving.It was noted in the service order that the device did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7797995
MDR Text Key117617222
Report Number8030965-2018-55822
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819977815
UDI-Public(01)7611819977815(11)151221
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Initial Date Manufacturer Received 08/06/2018
Initial Date FDA Received08/20/2018
Date Device Manufactured12/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-