• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/16/2018
Event Type  malfunction  
Manufacturer Narrative
Final analysis found burn marks to the ac power adapter which contributed to the field complaint of power up anomaly.
 
Event Description
It was reported that after a power outage, the transmitter was unable to be powered up.The power prongs were found to be charred.The transmitter was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7798611
MDR Text Key117694940
Report Number2938836-2018-09826
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/05/2018
Initial Date FDA Received08/20/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-