As of today, device return and additional information has been requested for this complaint but has not become available.Since neither the underlying medical documents nor device part details were received for investigation no thorough medical investigation, manufacturing record review and assessment of the reported event can be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this patient¿s particular case, our investigation remains inconclusive.No preventative or corrective action has been initiated as a result of this investigation.
|
It was reported that left hip revision surgery was performed.During the revision, the acetabular cup & femoral head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information has been requested for this complaint but has not become available.A review of the complaint history for the cup / head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup.Similar complaints have been identified for the head and this failure will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The medical documents were reviewed.Although the cobalt and chromium levels were reported to be 19.9 and 13.5 respectively, neither the unit of measure nor the specimen collection dates were reported.The reported discolored/murky fluid, osteolysis, and adverse local tissue responses are consistent with findings associated with metallosis; however, without the supporting lab results, imaging, and the analysis of the explanted components, the root cause of the reported reactions cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|