• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BIRMINGHAM HIP RESURFACING; BHR ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BIRMINGHAM HIP RESURFACING; BHR ACETABULAR CUP Back to Search Results
Device Problems Material Disintegration (1177); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Pain (1994); Therapeutic Effects, Unexpected (2099); Injury (2348)
Event Date 05/04/2015
Event Type  Injury  
Event Description
It was reported that left hip revision was performed due to pain, due to severe metal on metal with adverse local tissue response.
 
Manufacturer Narrative
As of today, device return and additional information has been requested for this complaint but has not become available.Since neither the underlying medical documents nor device part details were received for investigation no thorough medical investigation, manufacturing record review and assessment of the reported event can be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this patient¿s particular case, our investigation remains inconclusive.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the acetabular cup & femoral head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information has been requested for this complaint but has not become available.A review of the complaint history for the cup / head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup.Similar complaints have been identified for the head and this failure will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The medical documents were reviewed.Although the cobalt and chromium levels were reported to be 19.9 and 13.5 respectively, neither the unit of measure nor the specimen collection dates were reported.The reported discolored/murky fluid, osteolysis, and adverse local tissue responses are consistent with findings associated with metallosis; however, without the supporting lab results, imaging, and the analysis of the explanted components, the root cause of the reported reactions cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIRMINGHAM HIP RESURFACING
Type of Device
BHR ACETABULAR CUP
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa, warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key7800748
MDR Text Key117707991
Report Number3005975929-2018-00286
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received08/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FEMORAL HEAD, PART AND LOT NUMBER UNKNOWN; FEMORAL HEAD, PART AND LOT NUMBER UNKNOWN
Patient Outcome(s) Hospitalization; Required Intervention;
-
-