• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 90CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 90CM LENGTH; SCS LEAD Back to Search Results
Model Number 3189
Device Problem Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/18/2018
Event Type  Injury  
Manufacturer Narrative
In the event the device is returned to the manufacturer, the reported event cannot be analyzed via laboratory testing.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2 reference mfr.Report#: 1627487-2018-07737.It was reported that the patient's leads were auto-reducing, due to several impedances found to be out of range.In turn, the patient's leads were reprogrammed; but imaging was performed, indicating that the patient's leads have migrated.As a result, the patient may undergo surgical intervention at a later date.
 
Event Description
Device 1 of 2 reference mfr.Report#: 1627487-2018-07737.Follow up revealed that the patient's pns system was explanted and replaced, which addressed the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 90CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
michael mccane
9725268297
MDR Report Key7800909
MDR Text Key117720003
Report Number1627487-2018-07736
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734401715
UDI-Public05414734401715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/27/2019
Device Model Number3189
Device Lot Number6174508
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2018
Initial Date FDA Received08/21/2018
Supplement Dates Manufacturer Received10/26/2018
Supplement Dates FDA Received11/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL: 1194, SCS ANCHOR (X2); MODEL: 3662, SCS IPG
Patient Outcome(s) Other;
Patient Age46 YR
-
-