• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM AORTIC VALVE; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM AORTIC VALVE; REPLACEMENT HEART VALVE Back to Search Results
Model Number 8300AB27
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Arrhythmia (1721); No Code Available (3191)
Event Date 02/22/2018
Event Type  Injury  
Manufacturer Narrative
Peri-procedural arrhythmias, heart block, and other conduction disturbances are common in patients with underlying cardiovascular disease and/or conduction abnormalities.They can be exacerbated with standard perioperative medications, anesthesia, and/or instrumentation of the heart.Temporary pacemakers are inserted in all patients undergoing aortic valve replacement (avr).It is not uncommon for patients to have short term/reversible periods of heart block or arrhythmias following the procedure while their heart recovers from cardiopulmonary bypass.In many cases, the temporary pm is left in the patient for a short time following the procedure and then subsequently removed prior to discharge.Following surgical aortic valve replacement (avr), new-onset bundle branch block has been reported in 16% to 32% of patients and the need for permanent pacemakers in 3% to 8% of patients.The reason for post-operative av block after surgical avr is related to injury to the cardiac conduction system during surgical excision of the adjacent diseased valve and annular tissue.The close anatomical relationship between the aortic valvular complex and the branching atrioventricular bundle explains the possible development of conduction abnormalities following prosthetic aortic valve procedures.Arrhythmias and conduction disturbances are also common after surgical repair or replacement of the tricuspid and mitral valves due to the proximity of the conduction pathways.In this case, the patient required a permanent pacemaker on post-operative day two.Based on the available information, the root cause for the event remains indeterminable.However, it is likely that patient related factors and procedural factors contributed to the event.The device history record (dhr) is pending.A supplemental report will be submitted upon completion.
 
Event Description
Edwards received information via a clinical trial that a patient with a 27mm pericardial aortic valve developed sick sinus syndrome and a permanent pacemaker was implanted on pod #2 without incident.The outcome was noted as to be resolved.Patient was discharged home on pod #5.
 
Manufacturer Narrative
The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM AORTIC VALVE
Type of Device
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key7802044
MDR Text Key117769269
Report Number2015691-2018-03465
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150036  
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup,Followup
Report Date 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/22/2021
Device Model Number8300AB27
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/27/2018
Initial Date FDA Received08/21/2018
Supplement Dates Manufacturer Received08/30/2018
07/23/2020
Supplement Dates FDA Received09/06/2018
12/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
-
-