Event summary: visual inspection of balloon catheter, 2af283 / 67346, showed that the luer was broken.Smart chip verification indicated the catheter was used not.Without reprogramming the catheter, it was connected to the console and no system notices were received; catheter was recognized.The catheter passed electrical integrity test and the performance as per specification.In conclusion, the reported (guide wire lumen broken out of package) was confirmed through testing.The catheter failed the product returned inspection due to a broken luer.If information is provided in the future, a supplemental report will be issued.
|